TC5000 Series Inverted Microscope Instruction Manual

Page 2 - Table of Contents

The image below designates the main components of the TC5300 Series Phase Contrast Inverted Microscope. 1. Binocular Head, 30° Inclination 2. Eyep

Page 3

The image below designates the main components of the TC5500 Epi- Fluorescence Inverted Microscope. 1. Binocular Head, 30° Inclination 2. Eyepiece

Page 4 - 1.1 Microscope Features

2.1 Installation Site The microscope should be operated in a room with as little dust as practically possible. Keep your instrument away from solvent

Page 5

2.3 Microscope Set Up • As a first step, remove all components from the shipping container and remove the packing materials. Save the container and

Page 6

• If purchased, un-wrap the mechanical stage (MA380/05). Locate the two holes on the bottom side of the plain stage on the right hand side. The two t

Page 7 - 1.6 Warranty Notes

2.4 Installing the Phase Contrast Slider These instructions are only to be followed if you currently have a Model TC-5100, TC-5200, TC-5500 and TC-56

Page 8

2.5 Installing Fluorescence Option The following assembly steps are not required if fluorescence accessories have NOT been purchased. Prior to instal

Page 9

3. Next, remove the fluorescence access cover on the back of the microscope as shown in Figure 2a using a #2 Phillips screwdriver removing all seven

Page 10

12. When installed correctly, the fluorescence filter carriage will slide easily and click and lock in all three positions. When using the microsc

Page 11

2.6 Fluorescence Filter Information The epi-fluorescence equipped TC-5500 and TC-5600 include the following filter sets: Excitation Type Exciter Fi

Page 12 - Operating Ambient Conditions

Table of Contents 1.0 Introduction 1.1 Microscope Features 1.2 General Safety Guidelines 1.3 Warning/Caution symbols used in this manual 1.4 Int

Page 13 - 2.3 Microscope Set Up

3.0 Operation Once the microscope has been setup in its working location with all of the components correctly installed, it is ready for use. Your T

Page 14

3.1 Basic Set Up For Transmitted Light – Brightfield 1. First, set your interpupillary distance on the microscope eyetubes by pulling them apart or

Page 15 - Contrast

Objectives Information Depending on which objective is being used, before turning the nosepiece to change magnifications, the nosepiece will need to b

Page 16

3.2 Transmitted Light Operation – Phase Contrast Phase contrast is a very useful technique for high-contrast images of unstained or transparent spec

Page 17

Possible Phase Contrast Mode Operational Problems If normal adjustments are not getting the results you expect, check these: • Wrong phase annulus

Page 18

3.3 Incident Light Operation: Epi-Fluorescence For microscopes with fluorescence option installed only. For viewing of transparent objects using i

Page 19

Possible Incident Light Fluorescence Operational Problems If normal adjustments are not getting the results you expect, check these: • Weak fluores

Page 20

3.4 UV Light Safety Considerations Mercury arc lamps have UV light content in their output. The U.S. National Institute for Occupational Safety and

Page 21

3.5 Photomicrography with 35mm SLR and Digital SLR Cameras The TC Series of microscopes have a camera port directly in front of the instrument. In o

Page 22

3.6 Photomicrography with Digital Still Cameras In order to mount a consumer grade digital camera to this microscope, an optional camera attachment M

Page 23

TC-5000 Series Inverted Biological Microscope 1. Introduction The Meiji Techno TC-5000 Series inverted biological microscopes have a slim compact de

Page 24

4.0 Maintenance and Cleaning • Disconnect the power cord on your equipment prior to performing cleaning, maintenance or repair. • Keep electrica

Page 25

Cleaning of Glass Surfaces Use a soft brush or lint-free cotton cloth to removed dust and loose particles. For tough dirt, use a soft lint-free cott

Page 28

5.1 Replacing the mains fuse on the microscope ALWAYS DISCONNECT YOUR EQUIPMENT BEFORE DOING ANY REPAIR. Location of Mains Fuse The mains fuse of

Page 29

5.3 Replacing the 6V 30W Halogen Light ALWAYS DISCONNECT YOUR EQUIPMENT BEFORE DOING ANY REPAIR. Do not touch the glass envelope of the lamp durin

Page 30 - 4.0 Maintenance and Cleaning

5.4 Fluorescence Lamp does not work LAMP HOUSE MAY BE HOT TO THE TOUCH. Do not touch the glass envelope of the lamp at any time. Keep the protecti

Page 31 - Cleaning the Objectives

6.0 Storage Protect your microscope from dust after each use by covering your instrument with the protective dust cover that came with your microscop

Page 32

8.0 Accessories and Replacements Parts Part numbers and product descriptions for accessories and parts for TC Series Inverted Microscopes can be foun

Page 33

Epi-Fluorescent Objectives - Planachromat - Infinity Corrected - F = 200mm MA849 TC Planachromat F4X objective ∞/1.0, NA: 0.13, WD = 17.5mm, F.L.= 50.

Page 34

1.1 Microscope Features • CAD Designed Frame and Optics • Slim Compact Footprint • Infinity Corrected Optical System • Low Positioned Ergonomic C

Page 35

MA860 ND25 Neutral density filter for epi-illuminator (included with TC5500, TC5600)

Page 36

9.0 Technical Descriptions Filters and Applications Filter ND25 ND13 Green Interference LB100 G533 Cobalt Blue Application Ne

Page 37

10.0 Physical Dimensions

Page 38 - Eyepieces

Meiji Techno Co. Ltd. MEIJI TECHNO AMERICA 3010 Olcott St Santa Clara, CA 95054 Toll free: Phone: Fax: Email: Web: (800) 832-0060

Page 39

1.3 Warning/Caution Symbols Used in this Manual You must be aware of all safety issues when you install and operate this microscope system. Several w

Page 40

1.5 Product Safety Information- Handling the Microscope This microscope is not intended to be used in the immediate vicinity of w

Page 41 - Filters and Applications

1.6 Warranty Notes Meiji Techno warrants this product against defects in material and/or workmanship for the life of the instrument from the date of

Page 42

Meiji Techno’s convenience. In some cases, requests for warranty service are received which are not justified. In these cases, Meiji Techno will provi

Page 43

2.0 The Microscope and its Components The image below designates the main components of the TC5100 Series Brightfield Inverted Microscope. 1. Bino